OPTIMIZING SURFACE ELECTROMYOGRAPHY ACQUISITION WITHOUT RIGHT LEG DRIVE CIRCUIT

Authors

  • Zinvi Fu
  • A.Y. Bani Hashim Department of Robotics & Automation, Faculty of Manufacturing Engineering, Universiti Teknikal Malaysia Melaka, Hang Tuah Jaya, 76100 Durian Tunggal, Melaka, Malaysia
  • Z. Jamaludin Department of Robotics & Automation, Faculty of Manufacturing Engineering, Universiti Teknikal Malaysia Melaka, Hang Tuah Jaya, 76100 Durian Tunggal, Melaka, Malaysia
  • I. S. Mohamad Department. of Thermal-Fluid, Faculty of Mechanical Engineering, Universiti Teknikal Malaysia Melaka, Hang Tuah Jaya, 76100 Durian Tunggal, Melaka, Malaysia
  • N. Nasir Department of Robotics & Automation, Faculty of Manufacturing Engineering, Universiti Teknikal Malaysia Melaka, Hang Tuah Jaya, 76100 Durian Tunggal, Melaka, Malaysia

DOI:

https://doi.org/10.29121/IJOEST.v1.i1.2017.02

Keywords:

Circuit Design, Common Mode Noise, Electromyography, Right Leg Drive

Abstract

The right leg drive (RLD) is a circuit associated with electrocardiography acquisition circuits. For electromyography (EMG), the RLD circuit is used to a lesser degree. In general, the RLD circuit provides better noise reduction. This study compares the output of the EMG with and without the RLD circuit. The results indicate that with a good filter design, the direct grounding method can match the RLD in terms of noise reduction. As a result, EMG application, the RLD drive can be omitted.

Downloads

Download data is not yet available.

References

M. Guermandi, E. F. Scarselli, and R. Guerrieri, “A Driving Right Leg Circuit (DgRL) for Improved Common Mode Rejection in Bio-Potential Acquisition Systems,” IEEE Trans. Biomed. Circuits Syst., vol. 10, no. 2, pp. 507–517, Apr. 2016.

D. K. Freeman, R. D. Gatzke, G. Mallas, Y. Chen, and C. J. Brouse, “Saturation of the right-leg drive amplifier in low-voltage ECG monitors,” IEEE Trans. Biomed. Eng., vol. 62, no. 1, pp. 323–330, 2015.

M. M. Shobaki, N. A. Malik, S. Khan, A. Nurashikin, S. Haider, S. Larbani, A. Arshad, and R. Tasnim, “High Quality Acquisition of Surface Electromyography – Conditioning Circuit Design,” IOP Conf. Ser. Mater. Sci. Eng., vol. 53, p. 012027, 2013.

A. J. H. Mohideen and S. N. Sidek, “Development of EMG circuit to study the relationship between flexor digitorum superficialis muscle activity and hand grip strength,” 2011 4th Int. Conf. Mechatronics Integr. Eng. Ind. Soc. Dev. ICOM’11 - Conf. Proc., no. May, pp. 17–19, 2011.

J. Wang and J. E. Bronlund, “Surface EMG Signal Amplification and Filtering,” Int. J. Comput. Appl. (0975 – 8887), vol. 82, no. November, pp. 15–22, 2013.

H. Lee, K. Kim, and S. R. Oh, “Development of a wearable and dry sEMG electrode system for decoding of human hand configurations,” IEEE Int. Conf. Intell. Robot. Syst., pp. 746–750, 2012.

C. S. Wang, “A new AC-coupled amplifier for portable ECG without reference electrode,” Comput. Electr. Eng., vol. 39, no. 1, pp. 141–149, 2013.

C. J. De Luca, “Surface Electromyography : Detection and Recording,” 2002.

C. J. De Luca, L. Donald Gilmore, M. Kuznetsov, and S. H. Roy, “Filtering the surface EMG signal: Movement artifact and baseline noise contamination,” J. Biomech., vol. 43, no. 8, pp. 1573–1579, 2010.

P. Konrad, “The abc of emg,” A Pract. Introd. to Kinesiol., no. April, pp. 1–60, 2005.

Downloads

Published

2017-01-01

How to Cite

Zinvi Fu, Hashim, A. B., Jamaludin, Z., Mohamad, I. S., & Nasir, N. (2017). OPTIMIZING SURFACE ELECTROMYOGRAPHY ACQUISITION WITHOUT RIGHT LEG DRIVE CIRCUIT. International Journal of Engineering Science Technologies, 1(1), 13–19. https://doi.org/10.29121/IJOEST.v1.i1.2017.02