1.
INTRODUCTION
Many and
different kinds of mixtures, ranging in size from nanometers to several
hundreds of micrometers, are used in chemical and biotechnological processes.
Separation processes are commonly used to purify or extract these mixtures with
various physicochemical properties according to some components [1]. Most mixtures used in chemical and
biotechnological processes contain magnetic (para or ferro) particles [2]. The most obvious example of these particles
is magnetotactic bacteria and red blood cells [3], [4], [5], [6], [7] Non-magnetic bioparticles that do not show
magnetic properties can also be magnetized by adding artificial magnetic
particles (magnetite, chromium, nickel, etc.) [4], [8]. Conventional (classical) separation systems
are not sufficient to separate extremely small (
) particles. Even though these
particles are very small in size, they can be separated efficiently in magnetic
separators [5], [6], [8], [9], [10], [11], [12]. Matrix elements of magnetic separators,
especially ferromagnetic materials, can be easily magnetized using an external
magnetic field. Through this applied magnetic field, high gradient magnetic
fields are produced in the magnetized matrix regions (active regions) of
magnetic separators [13], [14], [15] , [16]. Magnetic bioparticles can be easily
separated by being exposed to the effect of great magnetic force if they are in
these regions [9], [17], [18]. Gradient magnetic fields are also widely
used in many areas of chemical engineering [19]. Advantages of gradient magnetic fields for
intensification of heat and mass transfer processes that occur especially in
catalytic reactors have been presented in the literature [20], [21], [22]. Furthermore, the effects of gradient
magnetic fields on the acceleration of chemical processes on water and other
chemical suspensions at the molecular level have also been presented in the
literature [23], [24], [25], [26], [27], [28], [29].
The
magnetic force acting on the magnetic bioparticles in the active regions of the
separator is calculated as follows;
(1)
Where,
is the magnetic susceptibility of the
particle,
is the vacuum permeability,
is the volume of the particle, and
is external magnetic field intensity. Since
this force (
) is many times greater than other
forces between bioparticles (Van der Waals, electrostatic, drift, etc.),
magnetic bioseparation can be performed effectively. In general, the magnetic
field force determined by Equation (1) should be kept as high as possible for effective
separation of bioparticles. As can be seen in Equation (1), the magnitude of
the magnetic force affecting bioparticles is directly proportional to the
magnetic field intensity (
) and field gradient (
). In bioseparation procedures,
considering that the magnetic susceptibility level of particles is around
[6] and magnetic field gradient is
around
, it is understood that the force
acting on the
particle in biologic systems is around
[1]. On the other hand, magnetic force at these
levels may be insufficient for the separation of bioparticles with weak
magnetic properties. In this case, it is possible to increase the strength of
the magnetic system by increasing the magnetic field gradient through the
creation of new structures. For this reason, an approximate empirical model of
the magnetic field gradient in the working areas of magnetic separators need to
be determined. These empirical formulas, should then be evaluated, compared,
and corrected with empirical coefficients according to data obtained from
magnetic separation experiments. In general, the generation and calculation of
magnetic field gradient is one of the most important fundamental problems of
the electromagnetic field theory. The results of many studies on this topic
have been presented in the literature [11], [17], [30]. Different methods are used for the
calculation of magnetic field gradients in Electromagnetic Field Theory [31]. However, the formulas obtained by methods
based on the calculation of the scalar magnetic field potential, commonly used
in classical electromagnetic field theory, are complicated and not appropriate
for the practical calculations in separation practice
[9], [11], [17], [32]. This prevents both the
progression of biomagnetic separation theory and the expansion of its
applicability in different areas. At the same time, the different geometries
used in various new structures of the developed magnetic separators, cause new
problems in the calculation and evaluation of magnetic field gradients. For
that reason, new simple solution methods need to be developed to calculate
gradient magnetic fields. For example, in recent years, gradient magnetic fields
of
or other types of permanent magnets, which
have different geometric structures, have been used in nano and micro fluidic
technological systems. However, the necessity
of using simple methods to calculate gradient magnetic fields remains
valid in such systems [33], [34].
In this
study, two main problems in terms of both theoretical and practical were
solved. First, a simple magnetic device or magnetometer was designed to
determine the magnetic field with high precision of about
And then, empirical expressions were produced
to determine the magnetic field or field gradients. For this purpose, using the
printer principle, a magnetometer with Hall
probes was designed and used. This device can measure the magnetic field in
three dimensions (
) automatically at precise
intervals
In the experiments, the gradient magnetic
fields generated by rectangular and disc-shaped
type magnets even electromagnets were
measured. At the same time, empirical models were created according to the
experimental results. The measurement results were demonstrated with tables and
graphics, and necessary comparisons, and recommendations have been presented.
2.
MATERIAL AND METHOD
Numerous
methods and setups are used to measure the gradient magnetic fields that
generated in narrow air gaps in magnetic systems [11], [17], [35]. The most practical and suitable of these
methods is the use of Hall probes [36]. Magnetic field intensity (
of uniform and symmetrical magnetic fields in
a small air gaps and the gradient of this field (
) can be easily measured with Hall probes [36]. However, there are many challenges in
measuring
or
between poles with different geometries
(sphere, cylinder, prism, etc.) because this magnetic fields are strongly
non-uniform fields. Therefore, in biochemical separation process, they show
different magnetic field effects at different points of the working regions.
Foremost among these challenges, is the safe fixed of the Hall probe in the
measurement area and the precise selection of measurement ranges. For these
non-symmetrical air gap geometries, it is especially important that the Hall
probe is safely fixed at different points for protected from vibrations and
oscillations, and measurement intervals can be adjusted automatically. Because
of the errors caused by failure to meet these conditions affect the measurement
results negatively. Developing new methods to eliminate such problems has
always been in the center of attention of researchers [36], [37]. In this study, we
described the details of the method to eliminate such problems.
2.1. Material
2.1.1. 3D PRINTER PRINCIPLES
MAGNETOMETER WITH HALL PROBES
The
magnetometer, which is an automatic measurement device that uses the Hall probe
and works with the
printer principle, has been designed and
tested to measure gradient magnetic fields in different air gaps in magnetic
systems [38]. This magnetometer shown in Figure 1a has a
programmable feature. Moreover, the Hall probe placed in the air gap between
the magnets can move independently from each other in the
directions. The Hall probe which is fixed, can
be moved directed independently in each direction with using three step motors.
The working area of the magnetometer, designed by us
produced in the TEKNOPARK that is part of Inonu University, is
, and its overall size is 
Figure
1: Magnetometer,
measuring system for the magnetic field; a)
magnetometer with axial Hall probe on measuring platform, b) magnetometer with
measuring cell.
The
magnetometer’s control system is supplied by a 12V DC voltage source. The
magnets can be placed on the platform shown in Figure 1a or measurement cell
shown in Figure 1b. Measuring cell has various apparatus for different magnets.
Due to the special design of the measuring cell, magnets with different
geometries can be placed in, and the distance between these magnets can be
adjusted (Figure 1b). The measuring cell also provides the measurement of
magnetic fields in small size solenoid and electromagnet. By modifying /
configuring the unit on which the Hall probe is attached for fixation, it is
also possible to determine the magnetic susceptibility of small-sized magnetic
materials or mixtures.
2.1.2. DESCRIPTION OF THE MATERIAL
In the
experiments,
type permanent magnets that have high magnetic
energy (
and most preferred in High Gradient Magnetic Separator
(HGMS) and Magnetic Drug Targeting (MDT) systems were used. Samples are
rectangular
magnet
dimensions and disc-shaped magnets with
diameter
, thickness
(Figure 2). Magnetic field intensity on the
surface of the magnets has been measured as
and
, respectively. Despite these
results, in the designed magnetometer, it is possible to easily measure the
magnetic fields created by magnets containing
,
type, and new type of hexaferrites and
composites based on them [39] used in magnetic systems. In the
magnetometer, it is possible to measure the fields of solenoids and similar
winding types of magnetic field sources with precision.

Figure 2: a) Rectangular and b) Disc-shaped
magnets.
2.1.3. MEASUREMENT
SYSTEM
GM05 type
Gaussmeter with Hall probe (HIRST-Magnetic Instrument Ltd-UK) was used in the
experiments (Figure 3). GM05 Gaussmeter has measurement and evaluation features
of AC magnetic fields in DC and
frequencies. Gaussmeter’s operating temperature
is
, and reproducibility is 5%. GM05
has two Hall probes,
long,
transverse, and
diameter axial (Figure 3). These probes with
an extension connection cable RS232 enables the measuring device to be kept far
enough from the magnetic field source and to protect the device from external
electromagnetic effects. When necessary, the magnetometer can be protected from
the influence of external magnetic fields with an outer shield such as a
multilayered
film [40]. In
principle, all similar Gaussmeters with Hall probes, like the measuring device,
can also be used in this system. Besides, measurement results can be obtained
more easily with the USB and RS232 connections in the new GM07 / GM08
Gaussmeters with GM modifications. The devices’ software is also Windows
compatible.

Figure 3: Magnetic field
measuring elements. GM05 type Gaussmeter, Hall probes (axial and transversal)
and permanent magnets.
2.2. THE METHOD
The
measuring points in the air gap are automatically selected according to the
preloaded program, and the measured values are recorded at the exact steady
state of the Hall probe. These measurement results can be recorded on the
computer connected to the magnetometer and displayed in tabular or graphical
format.
2.2.1. THE MATHEMATICAL FORMULATION OF THE PROBLEM
The
theory and practice of magnetic separation of bioparticles, especially MDT, are
generally like HGMS processes. Despite this, both MDT and biomagnetic
separation of different particles have many different properties. Unlike HGMS,
magnetic bioseparation or MDT have the following specific characteristics:
Gradient
magnetic fields are mainly generated by permanent magnets. Permanent magnets
can have different geometries (square, rectangle, cylindrical, disc, etc.) and
different dimensions.
In the
areas of separation or drug targeting, the maximum value of magnetic field
should be focused. For example, a cylindrical magnet should at least create
magnetic field in the sick areas (
). On the other hand, in some
experiments, it was determined that this distance would approximately be
in order to keep
sized iron-carbon particles inside the body [41]. Therefore, the generated magnetic
field force should have higher values in all points of working regions.
In the
MDT, the size of drug carrier particles is desired to be
. If the particles are larger than
these sizes, cause blockages in capillaries [42]. Therefore, it is desired that
the magnetically drug-attached carrier particles have a smaller size. On the
other hand, reduction in particle size causes the effect of the magnetic field
on these particles to weaken and the required blocking time in the affected
sick areas increases. While blocking time is
in the field of
area, blocking time can be
in the field of
area [42].
The particles
in bioseparation procedures could have solid, liquid, deformable plastic, and
other different physicochemical properties. On the other hand, suspensions,
emulsions, plasma, or highly concentrated medias used in magnetic separation
processes in chemical, medical, and biological environments may have
rheological properties. Therefore, it is necessary to consider the effect of
these properties of the carrier media on the separation process.
The size
of the magnets or sources of electromagnetic fields used in magnetic
bioseparation processes must be small, and the magnetic fields generated must
be large enough. Moreover, it is obvious that the energy and the magnetic field
intensity induced by permanent magnets are limited. Therefore, it is necessary to
know the changes of the magnetic field intensity
and gradient
in the target region of the magnetic field for
each separation process. This can be obtained from basic empirical and
theoretical models that are based on many experimental results for each state.
There are many theoretical models for the variation of the magnetic field
intensity generated by permanent magnets [43]. In micro and nano fluidic systems in
biotechnology, rectangular, disc and cylindrical shaped magnets are commonly
preferred [33], [43]. The characteristic dimensions of
rectangular and disc-shaped magnets are shown in Figure 2.
In terms
of practical application for bioseparation and drug targeting systems, the
component of the magnetic field on the axis perpendicular to the magnet surface
(e.g.,
in Figure 2) is more important. In general,
the determination of all components of the magnetic field generated by magnets
in practical engineering calculations is modeled by numerical methods [33], [41], [42]. In addition, approximate and useful
formulas for magnetic field calculation of magnets are also presented in the
literature [33], [43]. Some of these formulas are given below. If
is the magnetic field intensity at the center
of magnet
,
component can be determined as
follows.
for
rectangular magnets [43];
(2)
or
disc-shaped magnets [43];
(3)
for
cylindrical magnets [41];
(4)
for
magnetic circuit with parabolic shape confocal pole (MCPSP) [41];
(5)
Here
and
are constants, 
.
Although
Equation (2) and Equation (3), contain simple expressions, they may not be suitable
for theoretical investigations. Equation (4) and Equation (5) show that the
magnetic field intensity on the surface of the magnets (
) is infinite. This is
incompatible with reality for real magnetization systems. To simplify /
eliminate these issues, in this study, it is proposed to model the change of
magnetic field intensity on the
-axis with the following two
approximate formulas.
(6)
(7)
Here,
,
and
are experimentally determined correction
coefficients. According to the results of the experiments, the coefficients
and
included in Equation (6) and Equation (7) can
be determined for each magnet geometry. With the approach in this study, the
compatibility of the experimental measurements made in the
direction of rectangular and disc-shaped
magnets to the proposed models were evaluated.
3.
RESULT AND DISCUSSION
In the
experiments,
was measured according to surface coordinates
of the magnet using the field generated by the
magnets fixed on the measuring platform. In Figure 4, magnetic field
measurements were made on the surface of the rectangular shaped magnet in
symmetry
. At all levels (
and
), it has been observed that the
magnetic field values on the
are symmetrical. An increase in magnetic field
intensity was observed at the edges of the magnet [43]. In the experiments, 1178
measurements were made for different regions on the magnets. The results
obtained are shown in Figure 5, Figure 6, and Figure 7.

Figure
4: Magnetic field measurement results at the
surface of the rectangular magnet with respect to the symmetry of the
.
In the
practical applications of biotechnological systems, the magnetic force acting
on bioparticles, which are generally separated, is at a specific distance from
the magnet surface. Therefore, from the results obtained from measurements, the
direction of
was evaluated.
Figure 5: Magnetic field intensity changes of rectangular
magnet; a) from the center of
the magnet in the direction of the
, b)
in the
and
direction at the.
Figure 5
shows the graph of the magnetic field intensity changing as it moves away from
the center of the rectangular
magnet (
dimensions) in the direction of the
. As can be seen in Figure 5a, the
magnetic field intensity decreases rapidly as it moves away from the surface of
the magnet. This reduction becomes faster the further away from the surface.
However, it is seen that the decrease in the magnetic field is still at a
certain level, even at a sufficient distance from the magnet surface (
). At this distance, the magnetic
field has decreased from
to around
, that is, it has decreased by
approximately 64%. At
from the surface, this change is around 40%.
Similar variation is clearly seen in the
and
directions when
(Figure 5b). Unlike, the magnetic field of the
disc-shaped
magnet decreases faster as it moves away from
the center in the
direction. This change is shown in Figure 6
and Figure 7.

Figure
6: The graph of change in
on the surface
of the disc-shaped
magnet.
Figure 6
shows the measurement results of the changes in the
and
directions of the magnetic field intensity on
the surface of the disc-shaped
magnet
. As seen in Figure 6, the
magnetic field intensity has a clear symmetry. However, the further away from
the center of the disc
, this field intensity gets
greater. These similar change characteristics for permanent magnets are also
sufficiently explained in the literature [43].
Figure 7: Magnetic field intensity changes of
disc-shaped
magnet; a)
graph from the center of the
magnet in the direction of the
, b)
graph at the
.
As seen
in Figure 7a, the magnetic field intensity decreased from
to around
that is around 74% at a distance
from the surface of the magnet. This drop rate
in the disc-shaped magnet is approximately twice the drop rate of the
rectangular magnet. Similar changes of the magnetic field on the surfaces of
magnets have been observed on other coordinates (Figure 7b).
In
general, these changes are of great importance in bioseparation processes in
the gradient magnetic field. Consideration of this phenomenon is critical
especially in in-Vivo and in-Vitro applications of MDT systems.
Because the effectiveness of the MDT technique is determined by the level of
the magnetic field or field gradient applied to magnetic drugs [33]. As a result of these
evaluations, it is clearly seen that it is more advantageous to use rectangular
shaped magnets in magnetic separation processes. However, the effect of
magnetic field gradient may be more dominant for targeting magnetic drugs in
regions close to the surface of the magnet. In this case, it would be more
advantageous to use disc-shaped magnets. The magnetic field for each system
must be examined separately if magnets with different shapes are used in
biomagnetic separation and drug targeting systems. In other words, in systems,
both intensity and gradient change should be evaluated for the magnetic field
at different points of the working area by measuring. The three-dimensional
magnetic field measuring device, we covered in our study, is particularly important
in this respect.

Figure
8: The variation of the magnetic field intensity
along
for rectangular shaped
magnets.
In Figure
8 and Figure 9, the results of both experiments and calculations made according
to Equations (6) and (7) are given together for comparison. In Figure 8,
according to the formula in Equation (6), the changes of the magnetic field
intensity in the
-direction at different values of
the correction coefficient
are shown. As shown in Figure 8, the
measurement data and the data obtained from Equation (6) are similar in nature
for
. However, Equation (6) does not
conformity with the results of the experiments at
values. Also, as
increases, the unconformity increases rapidly.
Therefore, Equation (6) gives more accurate results for rectangular magnets
only at
.
However,
these graphics do not match in terms of the values. Although the change of the
magnetic field with the exponent law according to the
distance is theoretically convenient, both the
field intensity and the field gradient values calculated according to Equation
(6) are obtained with errors. According to this result, it is not advantageous
to use the magnetic field gradient generated by rectangular magnets in magnetic
separators.

Figure
9: Variation of the magnetic field intensity along
for disc-shaped magnets.
Alternatively,
it is clear from Figure 9 that the empirical formula proposed in Equation (7)
fits better with experimental data for disc-shaped magnets. The variation of
the magnetic field intensity in the
direction of the disc-shaped magnet
theoretically and experimentally coincides with each other. Moreover, the
correction coefficient
included in Equation (7) can be determined by
making fewer number (2 or 3) of experimental measurements for magnets with
different sizes. This clearly shows that the theoretical model given in
Equation (7) is more advantageous for practical calculations. In fact, the
performance of biomagnetic separators are determined by the magnetic field
gradient rather than magnetic field intensity. Given this situation,
, which is obtained from Equation
(6) and Equation (7) and defined as the force factor, is also useful to
evaluate. Figure 10 shows the graph of the force factor change in the
direction. As can be seen in Figure 10, under
equal conditions, the force factor (
) obtained from Equation (7) seems
to be more effective. It is more advantageous to use the theoretical model
determined by Equation (7) to evaluate the performance of these separators
during the design and use of biomagnetic separators.

Figure
10: The variation of the dimensionless magnetic
field gradient factor along the 
4.
CONCLUSIONS
The
following results are obtained from the studies conducted:
·
The use of gradient magnetic fields
remains an important method in developing and using new technologies.
·
To use gradient magnetic fields
effectively in different systems, the main magnetization system (magnet, coil,
electromagnetic, etc.) should be professionally designed and selected. The
magnetic field intensity and gradient of the magnetic system must be measured
and evaluated for each structure. In this case, some classical methods may not
be enough. Therefore, simpler magnetometers should be designed or created for
new structures, especially for smaller or narrower working regions.
·
Permanent magnet or magnetizing
elements used in these systems may exhibit different performance depending on
the structure of systems, the geometry of the working area and the
characteristics of the environment used. Measuring and modeling the magnetic
field and field gradient is required to correctly evaluate these properties
both theoretically and experimentally.
· To
evaluate the changes of the magnetic field and the field gradient,
printer principle designed measurement setup
(magnetometer) with Hall probes can be used. This magnetometer, which makes
accurate measurements in three directions, allows to adjust the measurement
steps less than
and to obtain the measurement data both in
tabular and graphical formats.
·
The specially designed measuring
cell assembly can evaluate measuring magnetic field values at different points
of magnetic elements of different dimensions and geometries.
·
Designed magnetometer can be used to
evaluate the magnetic field and gradient in special magnetic separators for
chemical and biomedical engineering processes.
· The
magnetic field measurement device, with Hall
probe can be used with the same success to measure and evaluate the magnetic
fields in different magnetophoresis processes for nano and micro fluidic
systems.
·
The simple empirical formula shown
in Equations (6) and (7) can be used to evaluate the experimental results,
magnetic field intensity, and magnetic field gradient.
·
The proposed methods will significantly
contribute to the theoretical and experimental determination of the performance
characteristics of magnetophoresis / magnetic separation processes, which have
wide application areas.
SOURCES
OF FUNDING
This research received no specific grant from
any funding agency in the public, commercial, or not-for-profit sectors.
CONFLICT
OF INTEREST
The
author have declared that no competing interests exist.
ACKNOWLEDGMENT
This study was supported by Inonu
University Scientific Research Projects Coordination Unit (BAP) within the
scope of the project numbered FDK-2017-775.
REFERENCES
[1] P. Van Hee, M. Hoeben, R. Van der Lans, and L. Van der Wielen,
"Strategy for selection of methods for separation of bioparticles from
particle mixtures," Biotechnology and Bioengineering, vol. 94, no. 4, pp.
689-709, 2006. [Online]. Available:
https://onlinelibrary.wiley.com/doi/pdfdirect/10.1002/bit.20885?download=true.
[2] A. L. Elrefai, T. Yoshida, and K. Enpuku, "Magnetic parameters
evaluation of magnetic nanoparticles for use in biomedical applications,"
Journal of Magnetism and Magnetic Materials, vol. 474, pp. 522-527, 2019.
[3] R. E. Dunin-Borkowski, M. R. McCartney, R. B. Frankel, D. A.
Bazylinski, M. Pósfai, and P. R. Buseck, "Magnetic microstructure of
magnetotactic bacteria by electron holography," Science, vol. 282, no.
5395, pp. 1868-1870, 1998. [Online]. Available:
https://science.sciencemag.org/content/sci/282/5395/1868.full.pdf.
[4] U. Lins and M. Farina, "Phosphorus-rich granules in uncultured
magnetotactic bacteria," FEMS microbiology letters, vol. 172, no. 1, pp.
23-28, 1999.
[5] D. Melville, F. Paul, and S. Roath, "Direct magnetic separation
of red cells from whole blood," Nature, vol. 255, no. 5511, pp. 706-706,
1975. [Online]. Available: https://www.nature.com/articles/255706a0.pdf.
[6] A. Bahaj, J. Watson, and D. Ellwood, "Determination of magnetic
susceptibility of loaded micro-organisms in bio-magnetic separation," IEEE
Transactions on Magnetics, vol. 25, no. 5, pp. 3809-3811, 1989.
[7] A. Bahaj, P. James, and F. Moeschler, "Efficiency enhancements
through the use of magnetic field gradient in orientation magnetic separation
for the removal of pollutants by magnetotactic bacteria," Separation
science and technology, vol. 37, no. 16, pp. 3661-3671, 2002.
[8] C. T. Yavuz, A. Prakash, J. Mayo, and V. L. Colvin, "Magnetic
separations: from steel plants to biotechnology," Chemical Engineering
Science, vol. 64, no. 10, pp. 2510-2521, 2009.
[9] M. Zborowski and J. J. Chalmers, Magnetic cell separation. Elsevier,
2011.
[10] C. Hoffmann, M. Franzreb, and W. Holl, "A novel high-gradient
magnetic separator (HGMS) design for biotech applications," IEEE
transactions on applied superconductivity, vol. 12, no. 1, pp. 963-966, 2002.
[11] V. Karmazin, "Theoretical assessment of technological potential
of magnetic and electrical separation," Magnetic and Electrical
Separation, vol. 8, 1970.
[12] V. Schaller et al., "Motion of nanometer sized magnetic
particles in a magnetic field gradient," Journal of Applied Physics, vol.
104, no. 9, p. 093918, 2008.
[13] J. Zeng, X. Tong, F. Yi, and L. Chen, "Selective capture of
magnetic wires to particles in high gradient magnetic separation,"
Minerals, vol. 9, no. 9, p. 509, 2019.
[14] H. Kang, J. Kim, H. Cho, and K.-H. Han, "Evaluation of Positive
and Negative Methods for Isolation of Circulating Tumor Cells by Lateral
Magnetophoresis," Micromachines, vol. 10, no. 6, p. 386, 2019.
[15] J. Lunacek et al., "Efficiency of high gradient magnetic
separation applied to micrometric magnetic particles," Separation Science
and Technology, vol. 50, no. 16, pp. 2606-2615, 2015.
[16] Z. Yildiz, T. Abbasov, and A. Sarimeseli, "Effect of the
Magnetization Properties of the Granular Beds and the Operating Parameters on
the Removal Ferrous Particles From the Waters by Using Magnetic Filter,"
Particulate Science and Technology, vol. 31, no. 2, pp. 109-113, 2013.
[17] J. Svoboda, Magnetic techniques for the treatment of materials.
Springer Science & Business Media, 2004.
[18] A. Khaligh and H. B. Ghavifekr, "Design of a MEMS-based
magnetophoresis micro-separator," in 2016 24th Iranian Conference on
Electrical Engineering (ICEE), 2016: IEEE, pp. 1495-1498.
[19] P. Hajiani and F. Larachi, "Ferrofluid applications in chemical
engineering," Int. Rev. Chem. Eng., vol. 1, pp. 221-237, 2009.
[20] M. Munteanu and F. Larachi, "Inhomogeneous magnetic field
effects on the hydrodynamic properties of multiphase catalytic reactors,"
International Review of Chemical Engineering, vol. 2, no. 1, pp. 150-154, 2010.
[21] M. Rolland, F. Larachi, and P. Hajiani, "Axial dispersion in
nanofluid Poiseuille flows stirred by magnetic nanoagitators," Industrial
& Engineering Chemistry Research, vol. 53, no. 14, pp. 6204-6210, 2014.
[22] F. Larachi, "Experimental and theoretical exploration of
weak-and strong-gradient magnetic fields in chemical multiphase
processes," Modeling of process intensification. Weinheim/Germany:
Wiley-VCH, 2007.
[23] V. Zablotskii, T. Polyakova, O. Lunov, and A. Dejneka, "How a
high-gradient magnetic field could affect cell life," Scientific reports,
vol. 6, no. 1, pp. 1-13, 2016.
[24] L. Mulay and I. L. Mulay, "Magnetic susceptibility:
instrumentation and analytical applications including bioscience, catalysis,
and amorphous materials," Analytical Chemistry, vol. 52, no. 5, pp.
199-214, 1980.
[25] C. L. Hill, A. Lamotte, W. Althoff, J.-C. Brunie, and G. M. Whitesides,
"High-gradient magnetic filtration of small particles of ferro-, ferri-,
and paramagnetic catalysts and catalyst supports," Journal of Catalysis,
vol. 43, no. 1-3, pp. 53-60, 1976.
[26] S. Ge et al., "Magnetic levitation in chemistry, materials
science, and biochemistry," Angewandte Chemie International Edition, vol.
59, no. 41, pp. 17810-17855, 2020.
[27] C. A. Sobecki, J. Zhang, and C. Wang, "Dynamics of a Pair of
Paramagnetic Janus Particles under a Uniform Magnetic Field and Simple Shear
Flow," Magnetochemistry, vol. 7, no. 1, 2021.
[28] J. Zhang and C. Wang, "Numerical study of lateral migration of
elliptical magnetic microparticles in microchannels in uniform magnetic
fields," Magnetochemistry, vol. 4, no. 1, p. 16, 2018.
[29] A. Eisenträger, D. Vella, and I. M. Griffiths, "Particle
Capture efficiency in a multi-wire model for high gradient magnetic
separation," Applied Physics Letters, vol. 105, no. 3, p. 033508, 2014.
[30] W. H. Hayt Jr, J. A. Buck, and M. J. Akhtar, Engineering
Electromagnetics| (SIE). McGraw-Hill Education, 2020.
[31] G. Lehner, Electromagnetic field theory for engineers and
physicists. Springer Science & Business Media, 2010.
[32] X. Zheng, Z. Xue, Y. Wang, G. Zhu, D. Lu, and X. Li, "Modeling
of particle capture in high gradient magnetic separation: A review," Powder
technology, vol. 352, pp. 159-169, 2019.
[33] E. P. Furlani, Permanent magnet and electromechanical devices:
materials, analysis, and applications. Academic press, 2001.
[34] K. M. Krishnan, Fundamentals and applications of magnetic materials.
Oxford University Press, 2016.
[35] S. Baik, D. Ha, R. Ko, and J. Kwon, "Magnetic field analysis of
high gradient magnetic separator via finite element analysis," Physica C:
Superconductivity, vol. 480, pp. 111-117, 2012.
[36] J. Sammer, M. Hubmann, F. Schauer, J. Schmidt, and J. Gerhold,
"Automatic measurement of nonuniform magnetic fields with industrial Hall
probes," IEEE transactions on magnetics, vol. 26, no. 5, pp. 2064-2066,
1990.
[37] S. A. Khashan, Y. Haik, and E. Elnajjar, "CFD simulation for
biomagnetic separation involving dilute suspensions," The Canadian Journal
of Chemical Engineering, vol. 90, no. 6, pp. 1450-1456, 2012.
[38] H. Bilgili, T. Abbasov, and Y. Baran, "Analysis and
Experimental Tests of Gradient Magnetic Fields for Separation of
Bioparticles," in II. International Battalgazi Multidisciplinary Studies
Congress, Malatya, Turkey, M. Talas, Ed., 15-17 March 2019 2019: IKSAD, 2019,
pp. 150-156.
[39] Y. Yang et al., "Influence of Nd-NbZn co-substitution on
structural, spectral and magnetic properties of M-type calcium-strontium
hexaferrites Ca0. 4Sr0. 6-xNdxFe12. 0-x (Nb0. 5Zn0. 5) xO19," Journal of
Alloys and Compounds, vol. 765, pp. 616-623, 2018.
[40] A. Trukhanov, S. Grabchikov, A. Solobai, D. Tishkevich, S.
Trukhanov, and E. Trukhanova, "AC and DC-shielding properties for the Ni80Fe20/Cu
film structures," Journal of Magnetism and Magnetic Materials, vol. 443,
pp. 142-148, 2017.
[41] O. Rotariu and N. J. Strachan, "Modelling magnetic carrier
particle targeting in the tumor microvasculature for cancer treatment,"
Journal of magnetism and magnetic materials, vol. 293, no. 1, pp. 639-646,
2005.
[42] L. E. Udrea, N. J. Strachan, V. Bădescu, and O. Rotariu,
"An in vitro study of magnetic particle targeting in small blood
vessels," Physics in Medicine & Biology, vol. 51, no. 19, p. 4869,
2006.
[43] Y. M. Pyatin, "Permanent magnets," Energiya, Moscow, 1980.